WRIGHTS LANE SYNTHES USA PRODUCTS LLC TI MATRIX CRVD SAGITL SPLIT PL/6H/12 BAR; PLATE, BONE
|
Back to Search Results |
|
| Model Number |
04.511.404 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Reporter is a j&j employee.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6), 2022 during orthognathic surgery, during the fixation of mandible following a bilateral sagittal split osteotomy procedure while inserting the screw into the plate, it protruded through the indicated plate hole, while attempting to secure the plate to the bone.This happened to the plate on the left and the right side of mandible during fixation.Procedure completed successfully.This report is for one (1) ti matrix crvd sagitl split pl/6h/12 bar.This is report 3 of 3 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: part: 04.511.404.Lot: l041521.Release to warehouse date: 30 june 2016.Expiration date : na.Supplier: na.Manufacturing site: werk mezzovico.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the ti matrix crvd sagitl split pl/6 h/12 bar (04.511.404) has bent in the plate holes area.The misalignment condition is attributable to device deformation.Also the plate holes has stripped.No other problems was reported.A dimensional inspection was not performed for the ti matrix crvd sagitl split pl/6 h/12 bar due to the design of the device.A functional test could not be performed as the mating devices were not returned.However, the misalignment issues could be confirmed due to the observed bent condition of the device.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the ti matrix crvd sagitl split pl/6 h/12 bar would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed current revision of drawing was reviewed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|