MAUDE Adverse Event Report: WELCH ALLYN INC. WIFI,CSM,NIBP,MASIMO,SURETEMP,PLUGB; MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WELCH ALLYN INC. WIFI,CSM,NIBP,MASIMO,SURETEMP,PLUGB; MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS) Back to Search Results
Model Number 75MT-B
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that the back housing was charred.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Additional Manufacturer Narrative
The connex spot monitors are intended to be used by clinicians and medically qualified personnel for monitoring of noninvasive blood pressure, pulse rate, noninvasive functional oxygen saturation of arteriolar hemoglobin (spo2), respiration rate, and body temperature in normal and axillary modes of neonatal, pediatric and adult patients.The most likely locations for patients to be monitored are general medical or surgical floors and general hospital and alternate care environments.This product is available for sale only upon the order of a physician or licensed health care professional.The 7000-ps is the connex spot monitor 35 watt power supply and the pwcd-2 is the power cord.The complete device was received at hillrom service center.It was found that the device would not power on during functional check.The problem was found to be with the main board burnt.All welch allyn electrical devices are designed and tested to meet all applicable safety and flammability regulations.Welch allyn's vital signs, cardio and monitoring electrical devices are not held by the user and therefore decrease the likelihood that the user would sustain a 1st or 2nd degree burn if the device were to become hot.If these burns were to occur, they would generally not be considered serious and would heal without further medical intervention.Information regarding the safety and warnings specific to each device is in the instruction for use (ifu).As per the ifu, warning patient injury risk.Damaged cords, cables, and accessories can affect patient and operator safety.Never lift the monitor by the power supply cord or patient connections.Routinely inspect the ac power cord, blood pressure cuff, spo2 cable, and other accessories for strain relief wear, fraying, or other damage.Replace as necessary.The ifu also states as part of a caution to replace the power cord if the strain relief, cord insulation, or metal prongs are damaged or begin to separate from the attachment plug.Additionally, if a device were to get hot to touch, spark, have burn marks, melted or burned components a trained clinician would not use the device and remove it from the patient's environment.There was no reported patient involvement or injury in this complaint, however, as this event could potentially lead to a serious injury or death if it were to recur, hillrom considers this complaint reportable.Hillrom is supplying the customer with a loaner device.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WIFI,CSM,NIBP,MASIMO,SURETEMP,PLUGB
Common Device Name
MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS)
Manufacturer (Section D)
WELCH ALLYN INC.
4341 state street
skaneateles NY 13153
Manufacturer Contact
kayla miller
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key14252092
Report Number1316463-2022-00056
Device Sequence Number5015892
Product Code MWI
UDI-Device Identifier00732094209372
UDI-Public732094209372
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K142356
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number75MT-B
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/01/2022
Initial Report FDA Received Date04/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-