Model Number 8637-20
Device Problems Filling Problem (1233); Migration or Expulsion of Device (1395)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/27/2022
Event Type  Injury  
Event Description
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a patient who was receiving morphine (20 mg/ml at 7.5 mg/day) via an implantable pump for spinal pain indications.It was reported that the healthcare provider (hcp) tried to fill the pump but was unable to do so because the pump had flipped.There were no external factors that contributed to the event.The hcp surgically access the pump pocket and flipped the pump to the correct orientation.They didn¿t aspirate the side port but confirmed the catheter was working and then tacked the pump back down.They then re-bolused the catheter postoperatively.The event was resolved at the time of the report.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key14252487
MDR Text Key290405660
Report Number3004209178-2022-05485
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00763000422608
UDI-Public00763000422608
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/28/2022
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/27/2022
Date Device Manufactured07/06/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age53 YR
Patient SexFemale
-
-