MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
|
Back to Search Results |
|
Model Number 8637-20 |
Device Problems
Infusion or Flow Problem (2964); Insufficient Information (3190)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/27/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: product id: 8780, serial# (b)(4), product type: catheter.
Other relevant device(s) are: ubd: 17-feb-2024, udi#: (b)(4).
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a patient who was receiv ing morphine (1.
5 mg/ml at.
1 mg/day) via an implantable pump for unknown indications for use.
It was reported that the physician was placing the spinal segment in intrathecal space and was unable to get csf (cerebrospinal fluid) back through the catheter after having perfect csf flow with needle.
The physician continuously moved the catheter but still was unsuccessful and asked for a second kit to be opened.
Turned out it was actually patient's anatomy.
Upon further review, it was found that the pump segment collette was broken.
There were no known factors that may have led or contributed to the issue.
A new catheter kit was opened.
The issue was resolved at the time of report.
|
|