MAKO SURGICAL CORP. 2.7 DEGREE ANGLED SAGITAL SAW; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 212480
Medical Device Problem Codes Application Program Problem: Parameter Calculation Error (1449); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
 
Event or Problem Description
Our distal cut was showing 1.5 deep with planar probe even though no red on screen was showing.We have had issues in last few weeks with cuts being too deep.We checked another cut made with straight attachment and it wasn¿t deep.Suspect that angled attachment may be going bad.No tourniquet was used.No surgical delay.Case type / application: tka.
 
Event or Problem Description
Our distal cut was showing 1.5 deep with planar probe even though no red on screen was showing.We have had issues in last few weeks with cuts being too deep.We checked another cut made with straight attachment and it wasn¿t deep.Suspect that angled attachment may be going bad.No tourniquet was used.No surgical delay.Case type / application: tka.
 
Additional Manufacturer Narrative
Reported event an event regarding inaccurate resection involving a mako saw attachment was reported.The event was not confirmed because the product was not available for inspection.Method & results -product evaluation and results: the product was not evaluated as the product was unavailable for inspection.If additional information is received, then the complaint will be reopened.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there are other complaints with a similar failure mode for this lot.Conclusions: the alleged failure mode was not confirmed because the product was not available for inspection.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7 DEGREE ANGLED SAGITAL SAW
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
diana rogers
210 centennial avenue
centennial park, elstree
borehamwood WD6 3-SJ
UK   WD6 3SJ
2082386500
MDR Report Key14253970
Report Number3005985723-2022-00051
Device Sequence Number17486943
Product Code OLO
UDI-Device Identifier00848486032128
UDI-Public00848486032128
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K143752
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number212480
Device Catalogue Number212480
Device Lot Number35030621 / 3518218
Was Device Available for Evaluation? No
Date Returned to Manufacturer04/18/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/07/2022
Supplement Date Received by Manufacturer12/30/2022
Initial Report FDA Received Date04/29/2022
Supplement Report FDA Received Date01/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/28/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-