| Model Number |
212480 |
| Medical Device Problem Codes |
Application Program Problem: Parameter Calculation Error (1449); Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
Our distal cut was showing 1.5 deep with planar probe even though no red on screen was showing.We have had issues in last few weeks with cuts being too deep.We checked another cut made with straight attachment and it wasn¿t deep.Suspect that angled attachment may be going bad.No tourniquet was used.No surgical delay.Case type / application: tka.
|
| |
|
Event or Problem Description
|
|
Our distal cut was showing 1.5 deep with planar probe even though no red on screen was showing.We have had issues in last few weeks with cuts being too deep.We checked another cut made with straight attachment and it wasn¿t deep.Suspect that angled attachment may be going bad.No tourniquet was used.No surgical delay.Case type / application: tka.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event an event regarding inaccurate resection involving a mako saw attachment was reported.The event was not confirmed because the product was not available for inspection.Method & results -product evaluation and results: the product was not evaluated as the product was unavailable for inspection.If additional information is received, then the complaint will be reopened.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there are other complaints with a similar failure mode for this lot.Conclusions: the alleged failure mode was not confirmed because the product was not available for inspection.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.
|
| |
|