| Model Number |
MICL12.6 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/23/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Unk.Investigation code (b)(6): lens work order search-no similar complaints reported for units within the same lot.Claim# (b)(64.
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 12.6mm micl12.6 implantable collamer lens, -9.5 diopter into the patient's eye.On (b)(6) 2022 the lens was removed due to excessive vault.Reportedly the lens was replaced.Attempts to obtain additional information have not been successful.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 - device evaluation: the lens was returned in liquid in an eye drop container.Visual inspection found no visible damage to the lens.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
B5 - the reporter indicated that the surgeon implanted a 12.6mm micl12.6 implantable collamer lens of -9.5 diopter into the patient's left eye (os) on (b)(6) 2022.On (b)(6) 2022, the lens was explanted.On the same date in a separate surgery the lens was replaced with a shorter length lens due to excessive vault.The problem is resolved.Cause of event is unknown.Claim# (b)(4).
|
| |
|