| Model Number |
117416 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Urinary Tract Infection (2120); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
04/05/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the survey respondent mentioned about bacteriuria, symptomatic urinary tract infection , catheter blockage , leakage when they were asked of the complications faced while using the temperature sensing foley catheter.It was unknown what medical intervention was provided.
|
| |
|
Event or Problem Description
|
|
It was reported that the survey respondent mentioned about bacteriuria, symptomatic urinary tract infection , catheter blockage , leakage when they were asked of the complications faced while using the temperature sensing foley catheter.It was unknown what medical intervention was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be ¿materials of construction are not biocompatible ¿.The lot number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.2.Using proper aseptic methods, remove catheter from package.3.Prepare patient per hospital/nursing recommended procedure.4.Proceed with catheterization using standard techniques.5.Inflate the 5cc balloon with a maximum of 10cc sterile water by inserting tip of water-filled luer-tip syringe gently into valve.Do not use a needle.6.Connect catheter to collection container.Connect temperature plug to appropriate patient monitor.7.To deflate the 5cc balloon, gently reinsert the luer tip syringe into the valve and aspirate.8.When temperature monitoring is no longer required, sever cord at valve arm with scissors and cover valve arm containing severed cord with adhesive tape." h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Search Alerts/Recalls
|