The device was not returned for analysis, as it was discarded by the customer; however,customer provided the picture of the sheath which showed that sheath was spilt at the distal end.Probable cause of the sheath split could be because the physician utilized the left groin for access, which is not the best practice or sheath met with the resistance while advancing into the patient body.No further follow-up is required.No corrective or preventive action resulted after investigation of this event.The event will be reevaluated if additional information becomes available.Oscor will continue to monitor this event type.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
|
|
The device was not returned for analysis, as it was discarded.The investigation was focused on a review of product documentation.Device was used in treatment.The device was not returned for analysis, therefore, the exact cause of the product issue cannot be determined and the clinical observation could not be confirmed.However, following controls are in place to mitigate the reported product issue.The complaint was confirmed using a photo provided by the customer.The device history records were reviewed to confirm that the device passed all applicable in-process and final inspections.Per procedure (abiomed introducer sheath in-process and final inspections): measure dimensions: measure tip i.D.Of sheath using appropriate pin gauges to be in range of 0.305" - 0.307" for 23f.Visual inspection: using 10x magnification, verify the tip is round, no flash protruding greater than 0.4 mm (0.016") and no flash with width (around circumference) greater than 0.7 mm (0.028"), no splitting, cracks or other damages.Per procedure, visual inspection, with naked eye at a distance of 12" to 18", verify the assembled sheath matches the drawing.Ensure parts are free of kinks, cracks, splits, sinks, excessive flash, loose/embedded fm, or any other damages.The instructions for use (ifu) informs the user: never advance or withdraw guidewire or sheath when resistance is met.Determine cause by fluoroscopy and take remedial action.When injecting or aspirating through the sheath, use the sideport only.Avoid subjecting the device to unusual stresses.When assembling the sheath and dilator, care must be taken to insert the dilator tip straight through the center of the valve membrane in order to prevent inadvertent puncturing of the membrane.No further follow-up is required.No corrective or preventive action resulted after investigation of this event.The event will be reevaluated if additional information becomes available.Oscor will continue to monitor this event type.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
|