| Model Number |
N/A |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
Perforation (2001)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Fujifilm healthcare corporation is performing additional investigation to determine the root cause.A supplemental report will be submitted pending results of the investigation.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2022, fujifilm healthcare corporation received a complaint regarding the arietta 850 ultrasound system.The site reported that during the therapeutic choledochoduodenostomy, the screen blacked out during the deployment of a boston scientific axios stent and a patient sustained a biliary perforation.A surgery was required to treat the perforation.The patient was stable post procedure.The event date is unknown.There is no death associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
On september 9, 2022, fujifilm healthcare corporation concluded the root cause investigation on the arietta 850.The error logs from the device were analyzed.It could not be confirmed that there was a causal relationship between the screen blacking out and the perforation.The available information was not enough to establish a root cause for the perforation.Fujifilm will continue to monitor all arietta 850 systems for similar issues.
|
| |
|