Model Number G30970
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/09/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter occupation: manager.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
As reported, an advance 18 lp low profile balloon catheter ruptured.The target location was the left superficial femoral artery.An inflation device was used to inflate the balloon twice before rupturing on the second inflation to nominal pressure.The device was removed from the patient, and the procedure was completed with a new balloon.There was no delay in treatment.It was initially reported that no stent was used during the procedure, but information was later received that the balloon was inflated inside of a stent prior to the rupture.Clarification has been requested.A section of the device did not remain inside the patient.The patient did not require any additional procedures due to the occurrence.According to the initial reporter, the patient did not experience any adverse effects.
 
Additional Manufacturer Narrative
Summary of event: as reported, an advance 18 lp low profile balloon catheter ruptured.The target location was the left superficial femoral artery.An inflation device was used to inflate the balloon twice before rupturing on the second inflation to nominal pressure.The device was removed from the patient, and the procedure was completed with a new balloon.There was no delay in treatment.It was initially reported that no stent was used during the procedure, but information was later received that the balloon was inflated inside of a stent prior to the rupture.A section of the device did not remain inside the patient.The patient did not require any additional procedures due to the occurrence.According to the initial reporter, the patient did not experience any adverse effects.Investigation evaluation: reviews of the complaint history, device history record, drawing, instructions for use (ifu), specifications, and quality control procedures were conducted during the investigation.The complaint device was not returned to cook for investigation.A document-based investigation evaluation was performed.No related non-conformances were found, and there have been no other reported complaints for this lot number.The product ifu warns ¿if balloon pressure is lost and/or balloon rupture occurs, deflate balloon and remove balloon and sheath as a unit.The information provided upon review of the dmr, dhr, and ifu suggests that there is evidence the device was manufactured to specification.There is no evidence of non-conforming devices in-house or in the field.Cook has concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Based on the information provided and the results of the investigation, cook has concluded that a component failure unrelated to manufacturing or design contributed to this incident.The risk analysis for this failure mode was reviewed and no additional escalation was required.The appropriate personnel have been notified and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information has been received since the last report was submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCE 18 LP LOW PROFILE BALLOON CATHETER
Common Device Name
LIT CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC.
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key14254858
Report Number1820334-2022-00672
Device Sequence Number14661018
Product Code LIT
UDI-Device Identifier10827002309705
UDI-Public(01)10827002309705(17)240201(10)13728755
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K130293
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/01/2024
Device Model NumberG30970
Device Catalogue NumberPTA4-18-150-7-20
Device Lot Number13728755
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/06/2022
Supplement Date Received by Manufacturer08/02/2022
Initial Report FDA Received Date04/29/2022
Supplement Report FDA Received Date08/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-