Model Number 1012068-09
Medical Device Problem Codes Product Quality Problem (1506); Difficult to Remove (1528); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: estimated.The device not returned to abbott vascular for analysis.A review of the electronic lot history record (elhr), corrective action tracking system for the web (catsweb) database review for this product was not performed since the lot number was not reported.There was no damage noted to the guide wire during the inspection prior to use which suggests a product quality issue with respect to manufacture, design or labeling did not contribute to the reported difficulties.Factors that may contribute to the difficulty to advance the guide wire may include, but not limited to, manufacturing, device placement technique, delivery catheter support, anatomical conditions/patient anatomical morphology, inner diameter of delivery devices, outer diameter of the guide wire, damage to the guide wire, damage to the delivery devices, coagulation of blood or procedural contaminates.The investigation was unable to determine a conclusive cause for the reported difficulty to advance, difficulty to remove and irregular texture.In this case, it is possible the guide wire encountered resistance with the anatomy causing the difficulty to advance and remove.The coils likely became damaged against resistance causing the irregular texture; however, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was performed to treat a lesion in the left superficial femoral artery (sfa).The versacore guide wire was advanced via crossover procedure, right to left, using a 6french sheath.During advancement, some resistance was noted and the device felt "funny".As the device didn't feel right, the decision was made to remove it.Some resistance was noted during removal and the device was removed intact.Upon removal, it was noted that the distal tip did not feel smooth, at the weld point.There was no adverse patient effect and no clinically significant delay in procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE VERSACORE MODIFIED J GUIDE WIRE SYSTEM
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14255888
Report Number2024168-2022-04697
Device Sequence Number20083691
Product Code DQX
UDI-Device Identifier08717648124235
UDI-Public08717648124235
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K083706
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1012068-09
Device Catalogue Number1012068-09
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/13/2022
Initial Report FDA Received Date04/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-