Model Number 410322-05
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/31/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) has requested that the vessel sealer instrument be returned for evaluation, but it has not yet been returned.Therefore, the root cause of the customer reported failure mode has not been determined.A follow-up mdr will be submitted if the product is returned and evaluated and/ or if additional information is received.A review of the site's complaint history does not show any additional complaints for this product.No image or procedure video was provided for review.A log review for the vessel sealer instrument could not be performed because the lot # was not provided by the customer.This complaint is being reported based on the following conclusion: it was reported that the vessel sealer instrument grips did not open while inside the patient.Medical intervention may be required in the event that the endowrist vessel sealer fails to unclamp from tissue when commanded by the user or system.While there was no harm or injury to the patient, the reported failure mode could likely cause or contribute to an adverse event if it were to recur.Blank mdr fields: follow-up was attempted, but the patient information was either unknown, unavailable, not provided, or not applicable.The expiration date is not applicable.Implant date is blank because the product is not implantable.Pma/510(k) number and adverse event are not applicable.
 
Event or Problem Description
It was reported that during a da vinci-assisted surgical procedure, the jaws of the vessel sealer instrument broke and did not open anymore.The procedure was completed with no reported injury.Intuitive surgical, inc.(isi) followed up with the initial reporter on (b)(6) 2022 and (b)(6) 2022, and obtained the following additional information: the doctor confirmed that the vessel sealer jaws were not stuck on tissue and the instrument was removed from the tissue as usual.No fragments fell inside that patient.A backup instrument was used to proceed with the surgical task.As a result of the alleged issue, there was a procedural delay of 80 minutes.There was no broken cable observed.
 
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) has received the vessel sealer instrument associated with this complaint and completed investigations.Failure analysis (fa) investigations did not replicate nor confirm the customer reported complaint.After retracting the blade, the vessel sealer instrument was placed and driven on an in-house system.The instrument passed the self-test homing.The instrument moved intuitively with full range of motion in all directions.The grips opened and closed properly.The instrument passed the knife slot test.An additional observation that was not reported by the site and was not related to the customer reported complaint was also identified.The vessel sealer instrument was found to have dislodged blade.Visual inspection showed that the blade was exposed outside of the blade garage 0.131".No conductor wire damage or snake wrist damage was found.There was no bio debris found at the instrument tip.A review of the logs showed an blade exposed failure.The root cause of the dislodged instrument blades is typically attributed to mishandling/ misuse.The vessel sealer instrument was found to have a self-test homing failure during in-house testing.Before retracting the blade, the instrument was placed on an in-house system.The instrument failed the self-test homing.This failure was a result of the dislodged blade failure of the instrument.The root cause of this failure is attributed to mishandling/misuse.A review of the instrument log for the vessel sealer instrument (part number 410322-05/lot number m10210726 494) associated with this event has been performed.Per logs, the vessel sealer instrument was last used in a procedure on 31-march-2022 on system (b)(6).
 
Event or Problem Description
Refer to h10/h11 for follow-up information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST;DAVINCI SI
Common Device Name
VESSEL SEALER
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key14265791
Report Number2955842-2022-11384
Device Sequence Number4438648
Product Code NAY
UDI-Device Identifier10886874111113
UDI-Public(01)10886874111113(10)M10210726
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K110639
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Pharmacist
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 04/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number410322-05
Device Catalogue Number410322
Device Lot NumberM10210726 494
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/01/2022
Supplement Date Received by Manufacturer06/24/2022
Initial Report FDA Received Date05/03/2022
Supplement Report FDA Received Date07/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DA VINCI INSTRUMENTS AND ACCESSORIES.
Patient Age75 YR
Patient SexMale
Patient Weight93 KG
-
-