MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SCR BIORCI 9X25MM STRL BX 1; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SCR BIORCI 9X25MM STRL BX 1; SCREW, FIXATION, BONE Back to Search Results
Model Number 7207563
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/09/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that, during an acl procedure, an 8.0 tunnel was made with a drill, the graft was used and a 1.5 nitinol guide was used, but at the time of admission the tip of the biorci screw broke.Surgery was resumed with a back-up device, but it is unknown whether it caused or not a surgical delay.No further complications were reported.
 
Additional Manufacturer Narrative
H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was no way to determine if the device contributed to the reported event.A complaint history review found no similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the resin found that the storage protocols, material specifications, and material tests were appropriately documented.A material certificate of analysis is required.It has not been reported whether the broken screw was retrieved from the patient.If a portion of the screw remains retained in the patient, it is made of a biocomposite material which is intended for implantation to allow for bone ingrowth.If the screw was retained it is unknown if the screw will migrate and there is potential risk for tissue inflammation and/or pain.No further clinical/medical assessment is warranted at this time.The complaint was not confirmed.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCR BIORCI 9X25MM STRL BX 1
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14266089
Report Number1219602-2022-00615
Device Sequence Number4439251
Product Code HWC
UDI-Device Identifier03596010426253
UDI-Public03596010426253
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K992396
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/06/2024
Device Model Number7207563
Device Catalogue Number7207563
Device Lot Number50844139
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/09/2022
Supplement Date Received by Manufacturer06/02/2022
Initial Report FDA Received Date05/03/2022
Supplement Report FDA Received Date06/06/2022
Date Device Manufactured12/06/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-