Model Number 01.18.462
Health Effect - Clinical Code Hip Fracture (2349)
Date of Event 04/01/2022
Type of Reportable Event Serious Injury
Event or Problem Description
At 1 month after the primary surgery, during a follow-up visit, stem sinking associated with peri-stem fractures is observed.However, there were no complaints from the patient such as pain, and she was able to walk without problems.The patient reject the revision surgery, but the surgeon will continue to recommend revision surgery.Fracture lines may be seen on x-rays post-primary.It is unclear whether the operation of the rasps during the surgery was bad, but it may be due to the fact that the rasps were taken in and out many times during the surgery due to unfamiliar surgery.Bone quality is unknown.
 
Additional Manufacturer Narrative
Batch review performed on 7 april 2022.Lot 2113027: 30 items manufactured and released on 15-feb-2022.Expiration date: 2027-feb-06.No anomalies found related to the problem.To date, 13 items of the same lot have been sold with no similar reported event.Clinical evaluation performed by clinical affairs specialist during a follow up visit few weeks after tha, it was noticed through radiographic images that the stem sinked and a fracture in the diaphyseal area was present.The sinking could have been caused by the fracture.The surgeon suggested a revision but since the patient did not report any pain she refused it.Intraoperative femoral fractures are known possible adverse events of total hip replacements, described and quantified in literature.They mainly depend on bone morphology and mechanical properties.In this case, there is no reason to suspect a malfunctioning device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEM: AMISTEM P AMISTEM- P SHORT NECK STD STEM SIZE 2
Common Device Name
CEMENTLESS HIP STEM
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key14267059
Report Number3005180920-2022-00327
Device Sequence Number5843098
Product Code LZO
UDI-Device Identifier07630971243668
UDI-Public07630971243668
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K192126
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number01.18.462
Device Catalogue Number01.18.462
Device Lot Number2113027
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/05/2022
Initial Report FDA Received Date05/03/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age72 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceAsian
-
-