| Model Number |
CREP G2 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
The initial reporter received a questionable crep2 creatinine plus ver.2 result for a patient sample tested on a cobas 6000 c 501 module with serial number (b)(4).The initial result was reported outside of the laboratory.On (b)(6) 2022, the initial result was 58.3 mg/dl.The first repeat result was 57.8 mg/dl with a data flag.The first repeat result was from an automatic rerun because of a middleware rule.The patient was sent to the emergency room (er) because they thought he had renal insufficiency.At the er, his creatinine result was reported to be at a low level.On (b)(6) 2022, the reporter reran the initial sample.The second repeat result was 48.6 mg/dl.The reporter also reran the same sample collected in a tube with a gel separator.The third repeat result was 8.6 mg/dl.
|
| |
|
Additional Manufacturer Narrative
|
|
A general reagent or hardware problem can be excluded because the provided calibration and qc data were within specifications.The reaction kinetics looked inconspicuous.The investigation did not identify a product problem.
|
| |
|
Search Alerts/Recalls
|