| Model Number |
102812-67A |
| Medical Device Problem Code |
Failure to Sense (1559)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
|
| |
|
Event or Problem Description
|
|
On 04 february 2022, senseonics was made aware of an incident where user received an early sensor replacement alert resulting in an early sensor removal.
|
| |
|
Additional Manufacturer Narrative
|
|
D2: product code corrected to qhj.
|
| |
|
Additional Manufacturer Narrative
|
|
The sensor was tested in-house, and the review of qc revealed a loss of chemical performance, which confirms the complaint.The investigation shows the system correctly disabled the sensor due to performance failure (low overall glucose signals leading to msp) and the system's self-test functions are working normally.Msp failures were investigated in (b)(4), which determined the root cause of these failures to be sensor's hydrogel oxidation.This sensor has the same failure mode and additional investigation is not necessary for this complaint.As part of resolution, the rma was authorized to offer the user a sensor replacement.H3.Device evaluated by manufacturer? yes.H6.Investigation findings updated to 3231.H6.Investigation conclusions updated to 4307.
|
| |
|