| Model Number |
3341 |
| Medical Device Problem Code |
Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Code |
Undesired Nerve Stimulation (1980)
|
| Date of Event |
03/21/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event is estimated.
|
| |
|
Event or Problem Description
|
|
Related manufacturer reference number: 1627487-2022-02171.It was reported the patient was experiencing uncomfortable stimulation when pressing on the ipg.Diagnostic testing showed high impedance on one contact.As a result, surgical intervention took place wherein the physician replaced the extensions to address the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event for inadequate shock/stimulation cannot be confirmed through products testing alone.Returned lead extension passed continuity and output resistance measurement.No root cause was identified for the loss of stimulation.
|
| |
|