• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL EXTENSION, DUAL 4 CHANNEL 10CM; SCS EXTENSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL EXTENSION, DUAL 4 CHANNEL 10CM; SCS EXTENSION Back to Search Results
Model Number 3341
Medical Device Problem Code Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Code Undesired Nerve Stimulation (1980)
Date of Event 03/21/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event is estimated.
 
Event or Problem Description
Related manufacturer reference number: 1627487-2022-02171.It was reported the patient was experiencing uncomfortable stimulation when pressing on the ipg.Diagnostic testing showed high impedance on one contact.As a result, surgical intervention took place wherein the physician replaced the extensions to address the issue.
 
Additional Manufacturer Narrative
The reported event for inadequate shock/stimulation cannot be confirmed through products testing alone.Returned lead extension passed continuity and output resistance measurement.No root cause was identified for the loss of stimulation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION, DUAL 4 CHANNEL 10CM
Common Device Name
SCS EXTENSION
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key14269348
Report Number1627487-2022-02172
Device Sequence Number4354593
Product Code LGW
UDI-Device Identifier05414734402279
UDI-Public05414734402279
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P010032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/30/2022
Device Model Number3341
Device Catalogue Number3341
Device Lot Number7363590
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/21/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/11/2022
Supplement Date Received by Manufacturer06/09/2022
Initial Report FDA Received Date05/03/2022
Supplement Report FDA Received Date06/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/31/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SCS IPG; SCS LEAD (X4)
Outcome Attributed to Adverse Event Other;
Patient SexFemale
Patient Weight77 KG
-
-