• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. FIBERTAK DOUBLE LOADED FW CL; FASTENER, FIXATION, NON-DEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. FIBERTAK DOUBLE LOADED FW CL; FASTENER, FIXATION, NON-DEGRADABLE, SOFT TISSUE Back to Search Results
Model Number FIBERTAK DOUBLE LOADED FW CL
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 04/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2022, it was reported by a sales representative via email that an ar-3600-2 fibertak suture anchor forked like tip broke inside the patient.When surgeon attempted to mallet in the anchor, the drill guide shifted and the tip of the fibertak inserter broke.The fragment were not retrieved and remain in the patients cortical bone.Surgeon created a new bone socket and used another anchor to complete the case.This was discovered during a bicep tenodesis procedure on (b)(6) 2022.No further issues has been reported.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIBERTAK DOUBLE LOADED FW CL
Common Device Name
FASTENER, FIXATION, NON-DEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key14269611
Report Number1220246-2022-04817
Device Sequence Number5158856
Product Code MBI
UDI-Device Identifier00888867206960
UDI-Public00888867206960
Combination Product (Y/N)N
Initial Reporter StateMT
Initial Reporter CountryUS
PMA/510(K) Number
K151230
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 05/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFIBERTAK DOUBLE LOADED FW CL
Device Catalogue NumberAR-3600-2
Device Lot Number14697359
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/12/2022
Initial Report FDA Received Date05/03/2022
Date Device Manufactured12/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-