• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. BIO-COMPRESSION SCREW 3 X 18MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. BIO-COMPRESSION SCREW 3 X 18MM; SCREW, FIXATION, BONE Back to Search Results
Model Number BIO-COMPRESSION SCREW 3 X 18MM
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 04/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that during an ocd repair surgery the bio-compression screws broke while inserting.The first one broke about 4mm to complete the insertion.The second one broke with half of the screw outside the bone.The tunnel was drilled with the correct drill size and tap.Apparently the screw was being introduced in the correct direction.The broken part remains in patient.The surgery was finished successfully with chondral darts ar-4005b-18.It was not necessary to switch the surgical technique or do a second surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIO-COMPRESSION SCREW 3 X 18MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key14269636
Report Number1220246-2022-04819
Device Sequence Number5158857
Product Code HWC
UDI-Device Identifier00888867034693
UDI-Public00888867034693
Combination Product (Y/N)N
Initial Reporter CountryPO
PMA/510(K) Number
K060478
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 05/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2023
Device Model NumberBIO-COMPRESSION SCREW 3 X 18MM
Device Catalogue NumberAR-5025B-18
Device Lot Number14231004
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/11/2022
Initial Report FDA Received Date05/03/2022
Date Device Manufactured09/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-