• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SENSEONICS INC. EVERSENSE TRANSMITTER; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SENSEONICS INC. EVERSENSE TRANSMITTER; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM Back to Search Results
Model Number 102208-500
Medical Device Problem Code Imprecision (1307)
Health Effect - Clinical Code Hypoglycemia (1912)
Date of Event 04/07/2022
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2022, senseonics was made aware of an incident where user experienced hypoglycemia due to inaccuracies in sensor readings.
 
Additional Manufacturer Narrative
The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
 
Additional Manufacturer Narrative
Based on the analysis, the sensor readings captured in the case notes was confirmed at the following time of the day on (b)(6).Sensor reading - 120 mg/dl at 2:02 am est on (b)(6) 2022.Sensor reading - 133 mg/dl at 2:02 am est on (b)(6) 2022.Per the case notes, the reported fingerstick measurement of 60-62 mg/dl was not entered in the system and since the exact time of the event is not captured, the mismatch event could not be confirmed at this time.The sensor readings leading up to the reported event, displayed very good agreement between the sensor readings and fingerstick measurements (refer to the attached screenshot).Per the analysis, the system was performing within expectations at the time.User was symptomatic, but did not specify if he sought any medical attention or self resolved the event.User was advised to let hcp know of these events.Per the investigation analysis, the system was performing within expectations at the time.No further resolution was possible for this complaint as user was unresponsive to the communications from customer care.H3.Device evaluated by manufacturer? yes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE TRANSMITTER
Common Device Name
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20875-7005
MDR Report Key14269713
Report Number3009862700-2022-00068
Device Sequence Number5908060
Product Code QHJ
UDI-Device Identifier00817491021441
UDI-Public00817491021441
Combination Product (Y/N)Y
Initial Reporter CountryUS
PMA/510(K) Number
P160048
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date04/12/2022
Device Model Number102208-500
Device Catalogue NumberFG-3300-01-001
Device Lot Number123496
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/08/2022
Supplement Date Received by Manufacturer04/08/2022
Initial Report FDA Received Date05/03/2022
Supplement Report FDA Received Date06/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/12/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age61 YR
Patient SexMale
-
-