| Model Number |
RELTACK3X10 |
| Medical Device Problem Codes |
Device Difficult to Setup or Prepare (1487); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, prior to a procedure, after removing the cartridge with 0 tack and after attempting to load a new cartridge, the handle and cartridge could not be attached.The nurse was able to press the red button and squeeze the handle to the end, and make the tip pin in the vertical position with the shaft mark, but the tip pin did not return to the position where the cartridge could be loaded, so it could not be loaded.After the loading issue occurred, the reload was used.A new tacker was taken out and the case was completed without any problems.There was no patient involved.
|
| |
|
Event or Problem Description
|
|
According to the reporter, prior to a procedure, after removing the cartridge with 0 tack and after attempting to load a new or second cartridge, the handle and cartridge could not be attached.The first cartridge was able to fire without any problem.The nurse was able to press the red button and squeeze the handle to the end, and make the tip pin in the vertical position with the shaft mark, but the tip pin did not return to the position where the cartridge could be loaded, so it could not be loaded.After the loading issue occurred, the reload was used.A new tacker was taken out and the case was completed without any problems.There was no patient involved.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted no abnormalities and no single use loading unit was received.Functionally, the articulation knob functioned properly.It was reported that there was difficulty in securing the reload on the device; however, the reload was still used for the procedure.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|