| Model Number |
180308HC-S-V-A2 |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that while embolizing a patient there was no microplex helical soft vta coil in the microplex system casing.No patient impact was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was stated to be available for evaluation but has not yet been returned.If the device is received the investigation will be performed and a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation of the returned coil system found the hypotube bent, the body coil kinked, and the implant not attached to the pusher.No indication using a detachment controller was found on the pusher's heater coil.However, further inspection found that the implant was stuck inside of the dispenser hoop.The investigation found the implant's monofilament with a tensile break shape at the tip which is consistent with the device experiencing excessive force that exceeded the strength of the monofilament causing the implant to separate from the pusher.The physical evaluation of the device could not identify the conditions or circumstances that led to the damage, but the damage is consistent with the device experiencing forces over specification.
|
| |
|