Model Number NM-401L-0425
Medical Device Problem Codes Mechanical Problem (1384); No Flow (2991)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported during a therapeutic procedure with nm-401l-0425 device, when sending liquid from a syringe, the liquid cannot be sent to the liquid sending port.The device was replaced.The intended procedure was completed with a similar device.There was no patient harm or injury reported due to the event.No user injury reported.
 
Additional Manufacturer Narrative
The subject device was received and evaluated.The device with model number nm-401l-0425 was returned for evaluation.The lot number was k1405.Device evaluation, the following were noted : the needle could extend from the outer tube.It was not possible to inject liquid when the slider was pushed.However, it was possible to inject liquid when the slider was pulled.The outer tube was not buckled.The needle tube presented compressive buckling.The dhrs (device history records) for this product have been reviewed.No abnormalities were detected in the device history record with the lot number for the following inspection items which related to the reported phenomenon.Process inspection sheet.Quality inspection sheet.Nonconforming product report.Ifu (instruction for use) review : the instruction manual contains the following descriptions, and it warns against this event.·straighten out the instrument before inspecting it.The instrument can be damaged if it is coiled while the handle is operated.·operate the slider slowly, otherwise the tube could buckle.·do not coil the insertion portion with a diameter of less than 15 cm.This could damage the insertion portion.·when inserting the instrument into the endoscope, retract the needle into the sheath, hold the instrument close to the biopsy valve, and keep it as straight as possible relative to the biopsy valve.Otherwise, the instrument could be damaged.Based on evaluation findings, it is determined that liquid could not be injected due to compressive buckling of the needle tube.Since friction resistance between the outer tube and needle tube has increased, compressive buckling might have occurred when the needle was extended.A likely mechanism causing friction resistance between outer tube and needle tube might be the following: the tube was coiled during the inspection for operation.The slider was pushed abruptly.Olympus will continue to monitor complaints for this device.
 
Additional Manufacturer Narrative
D8 - udi number was updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE INJECTOR
Common Device Name
INJECTOR AND SHEATHSET
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14270410
Report Number8010047-2022-07432
Device Sequence Number14878849
Product Code FBK
UDI-Device Identifier04953170422638
UDI-Public04953170422638
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K902736
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNM-401L-0425
Device Lot Number14K05
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/14/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/04/2022
Supplement Date Received by Manufacturer06/16/2022
Initial Report FDA Received Date05/03/2022
Supplement Report FDA Received Date07/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1