Model Number 1320-0310
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
It was reported that during u-lag screw extraction, the tip of the driver broke off.The surgery was completed with the preparation of a substitute.
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the device was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: the received u-blade lag screwdriver showed normal signs of usage.However, the pegs at the distal end of the outer sleeve were found to be broken which was due to the application of high torsional force.Closer view of the threaded rod we can see that the flanks threads on the proximal and distal area which might be due to heavy application of force can be due to hammering.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on the investigation, the root cause was attributed to being user-related.The appearance of a breakage surface indicates a brittle fracture caused most likely due to excessive torsional load.If any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
It was reported that during u-lag screw extraction, the tip of the driver broke off.The surgery was completed with the preparation of a substitute.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
U-BLADE LAG SCREWDRIVER GAMMA3®
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14271427
Report Number0009610622-2022-00164
Device Sequence Number12167611
Product Code HSB
UDI-Device Identifier04546540377845
UDI-Public04546540377845
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K032244
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1320-0310
Device Catalogue Number13200310
Device Lot NumberK136651
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/21/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/06/2022
Supplement Date Received by Manufacturer05/26/2022
Initial Report FDA Received Date05/03/2022
Supplement Report FDA Received Date06/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/01/2009
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age62 YR
Patient SexFemale
-
-