| Catalog Number |
2R8402 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a clearlink system non-dehp solution set was dripping and leaking below the drip chamber.It was not specified when in the process step this occurred.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(6).(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the device was received for evaluation.Visual inspection was performed using the naked eye which did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed including clear passage and pressure testing which revealed a leak between the assembly of the tubing into the drip chamber.The assembly of the tubing into the drip chamber was according to specification.Priming was performed on the set with sterile water which observed a leak at the affected area.Dimensional testing was performed at the tubing which met specifications.The reported condition was verified.The cause of the condition was due to an improper manual assembly during the manufacturing process.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|