Model Number ZA9003
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Eye Injury (1845)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that an intraocular lens (iol) was explanted from the patient's operative eye.The patient status was reported as temporarily impaired and daily activities were significantly affected.An incision enlargement and vitrectomy were not required.Through follow-up, the customer confirmed no further information is available.
 
Additional Manufacturer Narrative
Date of event: exact date unknown/not provided.Best estimate date is between 12/9/2021-3/31/2022.The device was not returned for analysis, therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.An attempt has been made to obtain missing information; however, the account did not have any other information.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Additional info: section d9: device available for evaluation? yes.Section d9: returned to manufacturer on: may 12, 2022.Section h3: device evaluated by manufacturer¿ yes.Device evaluation: the complaint lens was received in a different serial numbers lens case.Visual inspection under magnification revealed that the lens was received cut in half.The lens was cleaned and, no further issues could be observed with the lens that could contribute to an injury.Based on the return condition of the lens, no further product evaluation could be performed.The complaint issues were not confirmed.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Conclusion: as a result of the investigation there is no indication of a product quality deficiency, and the reported issue could not be confirmed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key14272618
Report Number3012236936-2022-01161
Device Sequence Number4355178
Product Code HQL
UDI-Device Identifier05050474529069
UDI-Public(01)05050474529069(17)261020
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
P990080
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberZA9003
Device Catalogue NumberZA90030225
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/07/2022
Supplement Date Received by Manufacturer06/23/2022
Initial Report FDA Received Date05/04/2022
Supplement Report FDA Received Date06/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight62 KG
Patient EthnicityHispanic
Patient RaceWhite
-
-