| Model Number |
10104-003 |
| Medical Device Problem Code |
Environmental Particulates (2930)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A field service engineer was dispatched to the customer site.The catalytic converter and oil mist filter were replaced to resolve the smoke/haze issue.Unit meets specifications and was returned to service.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Asp complaint ref #: (b)(4).
|
| |
|
Event or Problem Description
|
|
A customer reported an event of a ¿smoke¿ or haze emitting from the sterrad® 100nx sterilizer.There was no report of any injuries or human reactions.The customer powered off the unit and discontinued use.An asp field service engineer was dispatched to assess the unit onsite.This event is being reported as a malfunction report subsequent to a serious injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: asp investigation summary: the investigation included a review of the device history record (dhr), trending analysis of the smoke/haze issue, system risk analysis (sra), and functional analysis.¿ the device history record (dhr) was reviewed and no issues relating to the failure mode were noted.The involved unit met manufacturer specifications at the time of release.¿ trending analysis of the smoke/haze issue for the sterrad® 100nx unit was reviewed for the prior six months from open date and no significant trend was observed.¿ review of risk documentation shows the risk for exposure to toxic or corrosive material to be "low." the catalytic converter and oil mist filter were returned and successfully completed a test cycle with no evidence of smoke/haze or odor/smells issues.The reason for the failure and return of the catalytic converter and oil mist filter was not confirmed.The assignable cause of the smoke/haze is likely due to the catalytic converter and oil mist filter; however, the functional analysis could not replicate the reported issue.The asp field service engineer replaced the parts and confirmed the sterrad® 100nx was restored to proper function after service.The issue was resolved at the customer facility.Asp will continue to track and trend this issue.Asp complaint ref #: (b)(4).
|
| |
|
Search Alerts/Recalls
|