RESOLVE SURGICAL TECHNOLOGIES CABLE TENSIONER; INTERNAL ORTHOPAEDIC FIXATION SYSTEM, CERCLAGE WIRE/CABLE, NON-STERILE
|
Back to Search Results |
|
| Catalog Number |
391.201 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Device report from depuy synthes reports an event in colombia as follows: it was reported by the affiliate in colombia that during an unknown surgery on 4/6/2022, it was observed that the cerclage cable device burst.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.This report is for one (1) device.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: initial reporter occupation: reporter is a j&j employee.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Updated event description: it was reported that during a procedure on april 6, 2022, a diameter cable 1.0 was used.Initially it was tensioned up to 30 kg, but the cable was not tensioned enough and it was a few millimeters loose.The specialist asked to tighten the cable more, but when turning the cable tensioner up to 40kg, the cable burst.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.This report is for one (1) cable tensioner.This is report 1 of 2 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.A product investigation was conducted.The product was returned to depuy synthes for evaluation.The depuy synthes team forwarded the device to resolve surgical which conducted a visual inspection of the returned device.Visual analysis of the returned sample was performed on cable tensioner.Upon initial un-aided visual inspection it was found that this device does not have any major physical damage, however there are signs of wear from use.The tensioning knob seems to cycle completely in both directions with smooth action in both directions.Device still functions as intended.It passed all inspections.This device was then tested per supplier working instruction which is referred to in the routing step for assembly.Per this document, the device was tested with the following results: -36.6 kg -36.6 kg -36.2 kg.The dimensional inspection was not performed since dimensions are not relevant to complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed for the cable tensioner.The root cause of the complaint condition cannot be confirmed since no defect was detected.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.A device history record (dhr) review was conducted: part # 391.201 lot # p313874 manufacturing site: werk selzach logistik release to warehouse date: 23 may 2018 expiration date: n/a supplier: synthes usa hq, inc a manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|