|
The patient have had implanted an ika system.The spacer of the system seems to be rotated about 90°.It was diagnosed in (b)(6) 2017.The system was very well implanted and although it was rotated it worked sufficient so far that the surgeon decides to have a close follow-up and wait to revise the system.The patient have had no pain and walked well with the rotated system.In (b)(6) 2018 the surgeon decided to remove the system to avoid a permanent inappropriate load.During revision the nails have shown to be ingrown well so that the surgeon decided to just change the spacer.
|