SMITH & NEPHEW, INC. LENS 4K CCU WIFI VERSION; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 72205059
Medical Device Problem Codes Device Emits Odor (1425); Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/03/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an arthroscopy procedure, the lens 4k control unit started to smell after about 30 minutes and the system dropped out.The unit overheated/became hot to the touch, once the system cooled down, it run again.The case was completed with a non-significant delay using a back-up device.No patient complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference: case-(b)(4).The reported device was received for evaluation.A smith & nephew service center visual evaluation of the returned device reported the antenna was not returned.A smith & nephew service center functional evaluation of the returned device observed that the device turned off after 20-minutes.No mention of odor was reported.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.It was determined the device contributed to the reported event.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The complaint was confirmed and the root cause was associated with component failure.Factors which could have contributed to the reported event potentially include a mismatched or damp light cable.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future occurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LENS 4K CCU WIFI VERSION
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14274302
Report Number1643264-2022-00154
Device Sequence Number4354722
Product Code GCJ
UDI-Device Identifier00885556720844
UDI-Public00885556720844
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K191177
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Consumer,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72205059
Device Catalogue Number72205059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/28/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/08/2022
Supplement Date Received by Manufacturer06/03/2022
Initial Report FDA Received Date05/04/2022
Supplement Report FDA Received Date06/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1