| Model Number |
SRLP-205 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Codes |
Capsular Contracture (1761); Failure of Implant (1924)
|
| Date of Event |
04/05/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: device rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare provider reported a left side rupture.The device has been explanted and replaced.
|
| |
|
Event or Problem Description
|
|
Additionally, healthcare professional reported, capsular contracture, baker grade unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
The event of capsular contracture is a physiological complication.And analysis of the device generally does not assist allergan in determining a probable cause for this event.Reason for reoperation: capsular contracture, baker grade unknown.Device evaluation: analysis of the returned device identified, opening extended on anterior and underweight.A visual and microscopic analysis was performed which identified, opening crease sharp on anterior, crease fold, stress marks and wear abrasion.Based on the device analysis, the final assessment is: an opening crease sharp on anterior assessed, as fold flaw opening.
|
| |
|