• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH INC PEN NEEDLES FG; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH INC PEN NEEDLES FG; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Model Number PNDL, 5BV TVH TP 32G 4MM100CT 30/CS
Medical Device Problem Codes Fluid/Blood Leak (1250); Infusion or Flow Problem (2964)
Health Effect - Clinical Codes Pain (1994); Skin Inflammation/ Irritation (4545)
Date of Event 04/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal report reference number: (b)(4).Pen needles were not returned for evaluation.Note: manufacturer contacted customer in several follow-up calls to ensure the initial concern is resolved - unable to establish contact with customer at this time.
 
Event or Problem Description
Consumer reported complaint for the 32g trueplus pen needles.Customer stated that when he administers his insulin, the needle is really sharp, almost like there is a "burr" at the end of the needles.Customer stated it hurt when he sticks the needle into his skin.Customer stated the needle is moving around as he is administering the medication, and that he believed this was due to the needle not going in deep enough.Customer stated it was leaving welts underneath his skin from the fluid build up.Customer stated the fluid leaks out when he pulls the needle out.At the time of the call the customer feels well and did not report any symptoms.No medical intervention related to the use of the product was reported.The customer is not using compatible product.Coordinator had initially spoken with customer and transferred customer's information to technician.Technician was unable to contact the customer via telephone, no further information was able to be obtained.
 
Additional Manufacturer Narrative
Sections with additional information as of 16-jun-2022: h6: updated fda¿s type, findings and conclusions codes.H10: pen needles were not returned for evaluation.Complaint was forwarded to supplier quality and internal evaluation was performed by the manufacturer using pen needles from the same lot.No abnormalities observed on retain samples.Most likely underlying root cause: mlc-028: there was not enough information to determine the mlurc.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEN NEEDLES FG
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
TRIVIDIA HEALTH INC
2400 nw 55th court
fort lauderdale FL 33309
Manufacturer Contact
karen devincent
2400 nw 55th court
fort lauderdale, FL 33309
MDR Report Key14274423
Report Number1000113657-2022-00255
Device Sequence Number5908116
Product Code FMI
UDI-Device Identifier00021292010805
UDI-Public(01)00021292010805
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K180720
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 06/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/30/2023
Device Model NumberPNDL, 5BV TVH TP 32G 4MM100CT 30/CS
Device Catalogue NumberS6H01A32-100
Device Lot Number8G611
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Distributor Facility Aware Date04/04/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/04/2022
Supplement Date Received by Manufacturer05/20/2022
Initial Report FDA Received Date05/04/2022
Supplement Report FDA Received Date06/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-