| Model Number |
ETLW1616C124E |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/04/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product analysis conclusion: a type ib endoleak was confirmed on the films provided; however the cause of the event could not be determined.Lack of pre-implant ct's did not allow for a thorough assessment of the pre-implant anatomy and earlier post-implant films were not provided for comparison.It is most likely that disease progression, as noted by the physician, with iliac aneurysm morphology changes since implantation were a factor in the reported type ib endoleak.Analysis of the returned film did not reveal any out of specification stent graft integrity issues.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
An endurant ii limb stent graft was implanted during the endovascular treatment of an abdominal aortic aneurysm.It was reported that during a follow up ct , disease progression and a type ib endoleak of the right common iliac artery was found. the patient had a coil embolization treatment performed , with no issues.The physician plans to implant a limb extension at a later date.Per physician the cause was due to disease progression.No additional clinical sequelae were reported and the patient will be monitored.
|
| |
|