WILSON-COOK MEDICAL INC ACUSNARE POLYPECTOMY SNARE; KNS, ENDOSCOPIC ELECTROSURGICAL UNIT AND ACCESSORIES Back to Search Results
Model Number G22700
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
510k: k851958.Investigation evaluation: the product said to be involved was returned in an open pouch from the lot number provided in the report.The label matches the product returned.Our laboratory evaluation of the product said to be involved confirmed the report.The device was returned with the snare retracted beyond the distal tip of the catheter.During a visual inspection, the handle cannula was retracted and the drive wire was exposed in the handle slot.The hot pin was loose.A close visual examination of the proximal end of the handle cannula showed an indentation indicating the drive wire had been torqued.It is unknown how or when the hot pin became loose.A function test was not possible, due to the condition of the device.When the drive wire was manipulated manually with pliers, the snare head would advance and retract.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.Prior to distribution, all acusnare polypectomy snares are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered remote.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During a colonoscopy, the physician used a cook acusnare polypectomy snare.It was reported that wire was protruding from the [snare] handle.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACUSNARE POLYPECTOMY SNARE
Common Device Name
KNS, ENDOSCOPIC ELECTROSURGICAL UNIT AND ACCESSORIES
Manufacturer (Section D)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
MDR Report Key14277797
Report Number1037905-2022-00200
Device Sequence Number17232113
Product Code KNS
UDI-Device Identifier00827002227002
UDI-Public(01)00827002227002(17)250207(10)W4562945
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/07/2025
Device Model NumberG22700
Device Catalogue NumberASH-1-S
Device Lot NumberW4562945
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/05/2022
Initial Report FDA Received Date05/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ENDOSCOPE - OLYMPUS PCF-H190L
Patient SexUnknown
-
-