Model Number RLT281414
Medical Device Problem Codes Break (1069); Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to gore that a gore® excluder® aaa endoprosthesis (rlt281414) was intended to be used for an endovascular aneurysm repair (evar).The device was introduced in the aorta and placed under the renal arteries.Upon initial the deployment, resistance was experienced when pulling the handle.The deployment line was not visible and the physician believed that the deployment line broke inside the catheter very shortly after the handle.The physician tried to unfold the prosthesis via the backup deployment mechanism, however, it was not possible to pull the first deployment line further.The device was removed through the sheath still undeployed and a gore® excluder® conformable aaa endoprosthesis (cxt281412e) was implanted successfully.
 
Additional Manufacturer Narrative
No patient specific details have been provided.Therefore, the patient initials reflect the w.L.Gore internal case number.The following investigation is ongoing: product history review, engineering evaluation of the returned device.
 
Additional Manufacturer Narrative
A review of the manufacturing records indicated the lot met all pre-release specifications.The gore® excluder® aaa endoprosthesis has been returned to manufacturer.Engineering evaluation of the device showed the following: -visual examination showed that the backup hatch was removed, and the first deployment line was pulled but not completely.-engineering attempted to pull the first deployment line from the back up hatch and was unable to do so.The endoprosthesis and catheter bent as the line was pulled.-engineering took a closer look at the sleeve under magnification and the first two stitches of the sleeve were unlaced.The first deployment line appeared to be laced incorrectly at the third stitch, resulting in the line becoming stuck and unable to be deployed or unlaced.The physician¿s observation that the deployment line broke and the device was unable to be deployed from the backup deployment mechanism, was confirmed.The likely cause that the first deployment line broke and the device was unable to be deployed is the deployment line was laced incorrectly in the sleeve.Capa request was submitted to initiate review of this event.According to the gore® excluder® aaa endoprosthesis instructions for use (ifu), adverse events that may occur and / or require intervention include but are not limited to incomplete component deployment.Corrected h6: added health effect - impact code 4642.Added medical device problem code 3270.Added component code 788.Updated investigation findings code to 706 - assembly problem identified.Code 3233 is no longer applicable.Updated investigation conclusions codes to 23, 4315, and 22.Code 11 is no longer applicable.Code 706 refer to the type of investigation code 10 (testing of the returned device).Conclusions code 23: the likely cause that the first deployment line broke and the device was unable to be deployed is the deployment line was laced incorrectly in the sleeve.Conclusions code 4315: a review of the manufacturing records indicated the lot met all pre-release specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 2 B/P
32470 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
ida simson
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key14278143
Report Number3007284313-2022-01885
Device Sequence Number12090319
Product Code MIH
UDI-Device Identifier00733132622535
UDI-Public00733132622535
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/26/2024
Device Model NumberRLT281414
Device Catalogue NumberRLT281414
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/22/2022
Supplement Date Received by Manufacturer07/12/2022
Initial Report FDA Received Date05/04/2022
Supplement Report FDA Received Date07/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GORE® DRYSEAL FLEX INTRODUCER SHEATH 18F
Outcome Attributed to Adverse Event Other;
Patient Age86 YR
Patient SexMale
-
-