Model Number V11443-012B
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code Skin Burning Sensation (4540)
Date of Event 04/05/2022
Type of Reportable Event Malfunction
Event or Problem Description
We had another package open up and spill on employee¿s hand.No injury; however, it did get all over her hand and was pretty hot.She immediately washed the contents off.
 
Additional Manufacturer Narrative
Device history record review was completed on the reported lot number, v2c018.The lot was found to have been manufactured and released to predetermined specifications.No anomalies were found during review of the records.We received one sample for evaluation, and it was already activated.The root cause was determined to be torn pouch which resulted from improper machine set up.The responsible employee was retrained.A cross-functional team has been established to closely monitor burst complaints; the team is focused on identifying root causes of failure for samples returned to minimize reoccurrence.Cardinal health will continue to monitor for similar reports.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOVAPLUS HOT PACK, LARGE, 6X9
Common Device Name
PACK, HOT OR COLD, DISPOSABLE
Manufacturer (Section D)
MO016 MO-MOBERLY
808 w highway 24
moberly MO 65270
Manufacturer (Section G)
MO016 MO-MOBERLY
808 w highway 24
moberly MO 65270
Manufacturer Contact
patricia tucker
3651 birchwood drive
waukegan, IL 60085
8478874151
MDR Report Key14278539
Report Number1423537-2022-00762
Device Sequence Number6209540
Product Code IMD
UDI-Device Identifier10885380167218
UDI-Public10885380167218
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberV11443-012B
Device Catalogue NumberV11443-012B
Device Lot NumberV2C018
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/06/2022
Initial Report FDA Received Date05/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient Age35 YR
Patient SexFemale
Patient Weight70 KG
-
-