• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7590
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 85% stenosed target lesion was located in the severely tortuous and severely calcified mid left anterior descending artery.A 2.50mm x 20mm maverick balloon catheter was advanced for dilatation.However, during the second inflation at 12 atmospheres for 21-22 seconds, the balloon ruptured.The device was removed and the procedure was completed with another of same device.There were no patient complications reported and the patient condition was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a maverick 2mr balloon catheter.The device was visually and microscopically examined.There were numerous kinks to the hypotube of the device.There was contrast in the inflation lumen and balloon.The balloon was loosely folded.The device was prepped with an inflation device filled with water and connected to the inflation port to inflate the device.The device was able to inflate to rated burst pressure and deflate with no issue.Product analysis could not confirm reported event as the device inflated to rated burst pressure and deflated with no issue.
 
Event or Problem Description
It was reported that balloon rupture occurred.The 85% stenosed target lesion was located in the severely tortuous and severely calcified mid left anterior descending artery.A 2.50mm x 20mm maverick balloon catheter was advanced for dilatation.However, during the second inflation at 12 atmospheres for 21-22 seconds, the balloon ruptured.The device was removed and the procedure was completed with another of same device.There were no patient complications reported and the patient condition was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAVERICK
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14278986
Report Number2134265-2022-05109
Device Sequence Number20154641
Product Code LOX
UDI-Device Identifier08714729370185
UDI-Public08714729370185
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/20/2024
Device Model Number7590
Device Catalogue Number7590
Device Lot Number0027526945
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/28/2022
Supplement Date Received by Manufacturer06/24/2022
Initial Report FDA Received Date05/04/2022
Supplement Report FDA Received Date07/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexFemale
-
-