Model Number 4000-0107-01
Medical Device Problem Code Audible Prompt/Feedback Problem (4020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No product was returned.We are unable to confirm the reported complaint.If the product is returned, smiths medical will reopen this complaint for further investigation.
 
Event or Problem Description
It was reported that when the medfusion 4000 wireless pumps are in clinical mode and are in use in the departments, the pumps alarm after the infusion delivery is completed.The first alarm shows the message -system failure: primary audible alarm post-.Immediately after that, another a second alarm shows -system failure: acu watchdog failure-.This problem does not occur every time the pump is used, but has occurred intermittently over the past 3 years (2019-2022).Device was used to replicate the reported issue, however clinical engineering has found that the alarm does not occur in maintenance mode.As an example, from the pharmguard server, it was found that from (b)(6), the acu watchdog failure alarm had occurred a total of 58 times for 11 pumps.Clinical engineering was unable to duplicate problem when simulating clinical mode as well.The root cause is still not identified.Representative from smiths medical was contacted.Representative mentioned that they were aware of the problem and that one possible cause could be water seeping into the speaker box at the bottom of the pump.However, when examined, all pumps were found to be dry with no water seepage.More than 15 pumps are involved.No patient injury was reported.
 
Additional Manufacturer Narrative
Additional information was received indicating that the total number of events was 57 and not 58 as previously reported.
 
Additional Manufacturer Narrative
Additional information was received indicating that the total number of events was 57 and not 58 as previously reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDFUSION 4000 SYRINGE INFUSION PUMP
Common Device Name
PUMP, INFUSION
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key14280092
Report Number3012307300-2022-07491
Device Sequence Number4874613
Product Code FRN
UDI-Device Identifier15019517154313
UDI-Public15019517154313
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K111386
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4000-0107-01
Device Catalogue Number4000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/05/2022
Supplement Date Received by Manufacturer05/11/2022
05/11/2022
Initial Report FDA Received Date05/04/2022
Supplement Report FDA Received Date06/08/2022
06/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/24/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1