Model Number 101041265
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 03/25/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Customer service conducted troubleshooting including; verifying battery count; if the device powered on with the battery pack; and if the device powered on with the transformer.The pump powered up fine using the transformer.The customer sees no visible damage on the battery pack but the pump shuts off after 2 minutes when using the battery pack.The customer was first sent a new battery pack and then was sent a new breast pump.It was requested that she return her battery pack and pump to medela for testing and analysis.The customer was contacted by a complaint handler on multiple occasions, including in writing, to get additional information, with no response as of the date of this report.Based on the results of our internal investigation (reference number (b)(4)),it cannot be definitively concluded that the pump caused or contributed to the customer¿s mastitis.The estimated incidence of mastitis in lactating women, whether using a breast pump or not, according to published clinical literature can be as high as 33%.In fact, clinical guidelines suggest the use of a breast pump to facilitate withdrawal of breast milk during bouts of mastitis.The complaint rate of mastitis across all reported failures, across all medela breast pumps, is (b)(4)% for the period of january 2013 to august 2017.Mastitis is usually a benign, self-limiting infection with few consequences for the suckling infant.The risk of mastitis is higher among women who have breastfed previously, especially those with a history of mastitis.Riordan & wambach, 4th ed.P.294: breastfeeding and human lactation.However in this case, the mother¿s mastitis required prompt medical attention for pain relief and prescription antibiotics to avoid progression to overwhelming sepsis.
 
Event or Problem Description
On (b)(6) 2022, the customer alleged to medela llc that her pump in style max flow breast pump works with the power cord but doesn't power on with the battery pack.On 04/01/2022, she contacted us back.She has tried new batteries and still is having issues.On 04/15/2022, she contacted us again and is still having battery pack issues and now has mastitis and was prescribed an antibiotic.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP IN STYLE WARRANTY EN FR ES (YELLOWFIN)
Common Device Name
PUMP, BREAST, POWERED
Manufacturer (Section D)
MEDELA LLC
1101 corporate dr
mchenry IL 60050
Manufacturer (Section G)
MEDELA LLC
1101 corporate dr
mchenry IL 60050
Manufacturer Contact
dave kurudza
1101 corporate drive
mchenry, IL 60050
8004358316
MDR Report Key14280561
Report Number1419937-2022-00043
Device Sequence Number6528125
Product Code HGX
UDI-Device Identifier00020451412658
UDI-Public020451412658
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K200508
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number101041265
Device Catalogue Number101041265
Device Lot Number0000037233
Was Device Available for Evaluation? No
Distributor Facility Aware Date04/15/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/15/2022
Initial Report FDA Received Date05/04/2022
Date Device Manufactured12/11/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-