| Catalog Number |
195-100 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The consumer reported a conflicting result with the binaxnow covid-19 home test on (b)(6) 2022.The consumer took a test and got a negative result, after some minutes he started seeing a second pink line.Conformation test was not performed.No additional patient information, including treatment and outcome, was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.A supplemental report will be provided after completion.
|
| |
|
Additional Manufacturer Narrative
|
|
The supplemental report is being submitted to provide the investigation conclusion.Testing was performed at abbott diagnostics scarborough, inc.On retained kit lot 180167 with internal positive quality control samples and negative quality control swabs.All test results were valid and performed as expected.Additionally, the manufacturing records and quality control release testing was reviewed for test kit part number 195-100 / lot 180167 and device part number 195-430h / lot 178507.The lot met the required release specifications.A review of the complaints reported as false negative patient results (confirmed and unconfirmed, conflicting results) related to kit lot 180167 showed that the complaint rate is (b)(4) and a review of the complaints reported as false positive patient results (confirmed and unconfirmed, conflicting results) related to kit lot 180167 showed that the complaint rate is (b)(4).Abbott diagnostics scarborough was unable to determine the exact root cause of the reported issue; however, it could possibly be related to issues including the self-test user performance, interpretation of the result, or the specific patient sample.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data : h10: a review of the complaints reported as false negative patient results (confirmed and unconfirmed, conflicting results) related to kit lot 180167 showed that the complaint rate is (b)(4).A review of the complaints reported as false positive patient results (confirmed and unconfirmed, conflicting results) related to kit lot 180167 showed that the complaint rate is (b)(4).
|
| |
|