| Model Number |
SSL-290 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
04/07/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported "left side rupture." device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: underweight and opening.A microscopic analysis was performed which identify sharp edge and with non-penetrating nick assessed as surgical impact opening.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is: - a sharp edge opening in the posterior with stress marks due to surgical impact opening.- non-penetrating nick.Additional, changed, and/or corrected data: d9, h3, h6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "left side rupture." device has been explanted.
|
| |
|
Search Alerts/Recalls
|