• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) VERITY ADX XL; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) VERITY ADX XL; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number 5056
Medical Device Problem Code Pacemaker Found in Back-Up Mode (1440)
Health Effect - Clinical Codes Arrhythmia (1721); Dyspnea (1816); Syncope/Fainting (4411)
Date of Event 04/14/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that patient presented to emergency room after experiencing syncope and dyspnea.Upon interrogation it was found that patient's pacemaker was in backup mode.It was noted that the pacemaker was at elective replacement indicator.It was also noted from electrocardiogram that patient was in complete heart block.The pacemaker was explanted and replaced.The patient was stable.
 
Additional Manufacturer Narrative
The reported event of backup operation was confirmed.Analysis of the device image found that the device went into backup mode due to low battery voltage.After the device was restored from backup mode, the device was found to be at end of life (eol).Functional testing of device was normal.A longevity calculation found the battery depletion was normal based on the device usage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERITY ADX XL
Common Device Name
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key14281909
Report Number2017865-2022-09298
Device Sequence Number17823786
Product Code DXY
UDI-Device Identifier05414734006712
UDI-Public05414734006712
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P880086
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2017
Device Model Number5056
Device Catalogue Number5056
Device Lot NumberP000019410
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/14/2022
Supplement Date Received by Manufacturer06/14/2022
08/04/2022
Initial Report FDA Received Date05/04/2022
Supplement Report FDA Received Date06/14/2022
08/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/12/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-