| Model Number |
FLT-112S |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.We are currently unable to establish a relationship between the device and the issue reported.The device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported and a definitive root cause for the reported event could not be determined.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that during a fluent procedure on (b)(6), the out-flopak cracked during the initial prime cycle of the console.No harm to the patient or staff reported.No other information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Hologic has performed a health risk assessment based on the failure mode related to out-flopaks broken//damaged and the probability of occurrence for the possible harms was determined to be acceptable.No injuries, diseases or death have been reported, the failure mode is not likely to result in a death, serious injury or to cause adverse health consequences to the patient.Hologic has determined this event as not reportable and will cease reporting based only on this failure mode.Hologic will continue to monitor and trend this failure mode.In case of any serious injury or harm to the patient, hologic will trigger the resumption of mandatory reporting per 21 cfr 803.18.The last report submitted for this failure mode is recorded under 1222780-2022-00122.
|
| |
|
Search Alerts/Recalls
|