| Model Number |
CR2 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device unexpectedly shut down during patient use.This issue is patient related; however there was no adverse event reported.Upon evaluation of the customer's device, stryker observed that the device's lid magnet was missing from the lid.As a result, the device may not automatically power on when the lid is opened.This may lead to defibrillation therapy being delayed or unavailable, if needed.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker contacted the customer to request additional information on the patient.The customer informed stryker that no further patient information is available.Patient fields in which information is not provided were intentionally left blank.Stryker evaluated the customer's device and verified the reported issue.Stryker replaced the device's lid and battery to resolve the reported issue.Proper device operation was observed through functional and performance testing and the device returned to the customer for use.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device unexpectedly shut down during patient use.This issue is patient related; however there was no adverse event reported.Upon evaluation of the customer's device, stryker observed that the device's lid magnet was missing from the lid.As a result, the device may not automatically power on when the lid is opened.This may lead to defibrillation therapy being delayed or unavailable, if needed.
|
| |
|
Additional Manufacturer Narrative
|
|
Physical damage to the lid latch was determined to be the cause of the missing magnet.The depleted battery (from the missing magnet) was determined to be the likely cause of the unexpected device shutdown.An engineering investigation has determined that due to the design of the lpcr2 lid switch, absence of the lid magnet allows current to flow from the battery, even when the device is in standby mode.This will reduce the life of the battery.Therefore, the reported issue, the loss of lid/lid magnet, is associated with two hazardous situations: opening the lid does not turn on the device, and, higher than intended current draw while the device is in standby mode results in a prematurely depleted the battery.Replacement of the device lid so that the magnet is present in the device allows the lid switch to function as designed; turning on/off the device when the lid is opened and preventing current draw when the device lid is closed.The lpcr2 operating instructions has been updated to include additional troubleshooting tips to direct the customer to act when device behavior indicates the lid magnet may be missing.A new warning informs the customer of the risk associated with a missing magnet.
|
| |
|