MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION COYOTE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION COYOTE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 24706
Medical Device Problem Codes Difficult to Remove (1528); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that removal difficulties occurred.The patient presented with peripheral artery disease (pad).The 90% stenosed target lesion was located in the moderately calcified anterior tibial artery.Following atherectomy and pre-dilatation, residual stenosis was 30-40%.A 2.0mm x 220mm x 150cm coyote balloon catheter was advanced for dilation.The device was inflated at nominal pressure and was deflated without issue.However, it was noted that the balloon was difficult to remove but eventually removed without issue.The coyote balloon catheter was re-introduced but it would not even load on the wire.The device was removed and the procedure was completed with another of the same device.There were no patient complications and the patient was stable.
 
Event or Problem Description
It was reported that removal difficulties occurred.The patient presented with peripheral artery disease (pad).The 90% stenosed target lesion was located in the moderately calcified anterior tibial artery.Following atherectomy and pre-dilatation, residual stenosis was 30-40%.A 2.0mm x 220mm x 150cm coyote balloon catheter was advanced for dilation.The device was inflated at nominal pressure and was deflated without issue.However, it was noted that the balloon was difficult to remove but eventually removed without issue.The coyote balloon catheter was re-introduced but it would not even load on the wire.The device was removed and the procedure was completed with another of the same device.There were no patient complications and the patient was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr: returned product consisted of a coyote balloon catheter.The outer shaft, inner shaft, balloon, and tip were visually and microscopically examined.The batch lot was confirmed correct.Microscopic analysis confirmed kinks under the balloon 22.1 mm and 21.7 mm from the tip.When a guidewire was back loaded into the device it stopped 21.1 mm from the tip due to the kink.Removal of the guidewire had resistance.The tip showed damage consistent with use.Device analysis was able to confirm the reported complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COYOTE
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14282572
Report Number2134265-2022-04599
Device Sequence Number17823791
Product Code DQY
UDI-Device Identifier08714729796466
UDI-Public08714729796466
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K111295
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/10/2025
Device Model Number24706
Device Catalogue Number24706
Device Lot Number0028661311
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/11/2022
Supplement Date Received by Manufacturer05/20/2022
Initial Report FDA Received Date05/04/2022
Supplement Report FDA Received Date06/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-