It was reported that during the implant attempt, the physician connected the right ventricular (rv) lead and left ventricular (lv) leads to the device with no issues.When inserting the wrench into the atrial port, the wrench was not engaging the setscrew.The setscrew would not loosen or tighten.The physician stated that it seemed like the wrench never fit into the head of the screw.A second wrench was tried with the same results on the atrial port.The physician then requested a new device.The new device was implanted successfully.No patient complications have been reported as a result of this event.
|
Product event summary: the device was returned and analyzed.Returned product analysis was performed.And no anomalies were found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank, because the information is currently, unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|