ETHICON INC. SFX SPI PDS+ UNI VIO 27IN 2-0 S/A CT-2; SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE Back to Search Results
Model Number SXPP1B414
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent a spine procedure (b)(6) 2022 and barbed suture was used.During the procedure, multiple sutures had 'needles breaking in half and ends not holding.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).(b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained via: fixation tab or loop breakage during the procedure or barbs are missing and not holding in the tissue during the procedure? loop was broken.Do you have a picture for visual analysis? yes have attached a picture.Was needle removed from the patient? needle did not go in the patient.Any additional tissue dissection damage due to searching for needle? no, the needle did not break inside the patient.It broke when they grabbed it with the needle driver.What was the size of the needle used? ct-2.Was more than half the needle retained in the patient? no needle in patient lot number? no.Trade name - irgacare®, active ingredient(s) ¿ triclosan, dosage form ¿ suture/solid/parenteral, strength ¿ = 2360 ug/m.Events reported via: 2210968-2022-03280 and 2210968-2022-03144.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 05/25/2022.H6 component code: g07002 - device not returned.This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803.Part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Photo analysis: visual analysis of the individual image submitted for evaluation determined that a section of suture could be seen.The image is not clear to determine failure mode or reported condition.As with any device, care must be taken to avoid damaging the thread when handling it.Avoid crushing or crimping action of surgical instruments such as needle holders and forceps.No conclusion could be reached as to the cause of the reported complaint, as the sample was not returned for analysis.As part of quality process, all devices are manufactured, inspected and released to approved specifications.Additional information about complaints is tracked for potential safety signals through complaint trends as part of post-market surveillance.The manufacturing records could not be reviewed as the batch number is unknown.Trade name - irgacare®.Active ingredient(s) ¿ triclosan.Dosage form ¿ suture/solid/parenteral.Strength ¿ = 2360 g/m.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SFX SPI PDS+ UNI VIO 27IN 2-0 S/A CT-2
Common Device Name
SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km. 8.3
san lorenzo PR
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
MDR Report Key14283674
Report Number2210968-2022-03279
Device Sequence Number5283444
Product Code NEW
UDI-Device Identifier10705031235878
UDI-Public10705031235878
Combination Product (Y/N)Y
Initial Reporter CountryUS
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSXPP1B414
Device Catalogue NumberSXPP1B414
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/05/2022
Supplement Date Received by Manufacturer05/04/2022
Initial Report FDA Received Date05/04/2022
Supplement Report FDA Received Date05/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1