Model Number 72200873
Device Problems Overheating of Device (1437); Electrical Shorting (2926)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/11/2022
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that during a knee arthroscopy, the dii controller had a short-circuited, it grew very hot with a burning smell.The procedure was completed with a s+n back-up device.Non-significant delay was reported, and no patient complications were reported.
 
Manufacturer Narrative
H3, h6: the reported device was received for evaluation.It was determined the device contributed to the reported event.A visual inspection found no issues.A functional evaluation revealed a short circuit detected error message in port a, the resistor associated with this port was burned out.No overheating was noted.An unable to retrieve custom settings error message caused by a low battery voltage was also found.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The complaint of overheating was not confirmed.Factors that could have contributed to the reported event include a failure of a concomitant device.The complaint of a short-circuit was confirmed, and the root cause has been associated with an electrical component failure.Factors that could have contributed to the reported event include connecting a hand piece that is still wet from sterilization processing or connecting a shorted out hand piece causing damage to the motor controller pcb.Please refer to the dyonics power ii control system operations/service manual for recommendations on proper cleaning and sterilization processes the complaint of a unable to retrieve custom settings was confirmed, and the root cause has been associated with an electrical component failure.Factors, unrelated to the manufacturing and design of the device which could have contributed to the reported event, include a low or dead battery voltage for the custom setting memory chip.This would cause a loss of power to main pcb's ic with the custom settings memory.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DII CONTROLLER
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
MDR Report Key14283685
MDR Text Key290792235
Report Number1643264-2022-00160
Device Sequence Number1
Product Code HRX
UDI-Device Identifier03596010607409
UDI-Public03596010607409
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062849
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72200873
Device Catalogue Number72200873
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/21/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/15/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-