• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ARTHROPIERCE INSTRUMENT CURVED LEFT; ACCESSORIES,ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ARTHROPIERCE INSTRUMENT CURVED LEFT; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Model Number 72201186
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 04/13/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during an arthroscopy, the arthropiece clamp was damaged at the time of passing the points of the second suturefix ultra 1.9mm anchor.The procedure was completed with a non-significant delay using a back-up device.The tip of the clamp remained inside the patient and could not be removed.Patient condition is stable.No further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found no similar reported events.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.This case reports the breakage of the arthropierce clamp, per case details the tip of the clamp remained inside the patient and could not be removed.The arthropierce clamp is made of 17-4 ph stainless steel rc 40-48.The clamp is an externally communicating device, it is neither manufactured nor intended for implantation.It is also not approved for long term internal tissue exposure and long-term implantation data is not available.The patient impact beyond the reported device breakage and retained foreign body could not be definitively determined.Mico-motion or movement cannot be predicted.There is potential risk for tissue inflammation and/or pain.No further clinical/medical assessment is warranted at this time.There was no way to determine if the device contributed to the reported event.The complaint was not confirmed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTHROPIERCE INSTRUMENT CURVED LEFT
Common Device Name
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14283742
Report Number1219602-2022-00633
Device Sequence Number12314489
Product Code NBH
UDI-Device Identifier03596010605610
UDI-Public03596010605610
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72201186
Device Catalogue Number72201186
Device Lot Number50581680
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/13/2022
Supplement Date Received by Manufacturer07/12/2022
Initial Report FDA Received Date05/04/2022
Supplement Report FDA Received Date07/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/07/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexMale
Patient Weight62 KG
-
-