BOSTON SCIENTIFIC CORPORATION CHARGER; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 24681
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The stenosed target lesion was located in the iliac vein.A 12.0 x80, 75cm charger balloon catheter was advanced for dilatation.However, during inflation at 10 atmospheres, the balloon ruptured.The procedure was completed with another of the same device.There were no patient complications reported and the patient status was stable.
 
Additional Manufacturer Narrative
Updated b5: describe event or problem.
 
Event or Problem Description
It was reported that balloon rupture occurred.The stenosed target lesion was located in the iliac vein.A 12.0 x80, 75cm charger balloon catheter was advanced for dilatation.However, during inflation at 10 atmospheres, the balloon ruptured.The procedure was completed with another of the same device.There were no patient complications reported and the patient status was stable.It was further reported that the balloon was inflated once for a few seconds before it ruptured.
 
Event or Problem Description
It was reported that balloon rupture occurred.The stenosed target lesion was located in the iliac vein.A 12.0 x80, 75cm charger balloon catheter was advanced for dilatation.However, during inflation at 10 atmospheres, the balloon ruptured.The procedure was completed with another of the same device.There were no patient complications reported and the patient status was stable.It was further reported that the balloon was inflated once for a few seconds before it ruptured.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHARGER
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14283777
Report Number2134265-2022-05178
Device Sequence Number6256272
Product Code FGE
UDI-Device Identifier08714729803881
UDI-Public08714729803881
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K112697
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/28/2024
Device Model Number24681
Device Catalogue Number24681
Device Lot Number0028290086
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/27/2022
Supplement Date Received by Manufacturer05/06/2022
07/13/2022
Initial Report FDA Received Date05/04/2022
Supplement Report FDA Received Date05/18/2022
07/25/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age66 YR
Patient SexMale
-
-